Validation of a Highly Sensitive RP-HPLC Method for Quantification of Fenofibrate in Pure and Pharmaceutical Dosage Forms

ISSN: 1875-676X (Online)
ISSN: 1573-4129 (Print)


Volume 10, 4 Issues, 2014


Download PDF Flyer




Current Pharmaceutical Analysis

Aims & ScopeAbstracted/Indexed in


Submit Abstracts Online Submit Manuscripts Online

Editor-in-Chief:
Atta-ur-Rahman, FRS
Honorary Life Fellow
Kings College
University of Cambridge
Cambridge
UK
Email: cpa@benthamscience.org

View Full Editorial Board

Subscribe Purchase Articles Order Reprints

Current: 0.771
5 - Year: 1.107

Validation of a Highly Sensitive RP-HPLC Method for Quantification of Fenofibrate in Pure and Pharmaceutical Dosage Forms

Author(s): Abid Mehmood Yousaf, Dong Wuk Kim, Han-Gon Choi and Euichaul Oh

Affiliation: College of Pharmacy & Institute of Pharmaceutical Science and Technology, Hanyang University, 1271, Sa-3-Dong, Ansan 426-791, South Korea.

Abstract

A simple, sensitive, specific, robust, precise and accurate reverse-phase high performance liquid chromatographic (RP-HPLC) method was developed and validated for suitable quantitation of fenofibrate in bulk and pharmaceutical dosage forms. The investigation of various validation parameters such as system suitability, linearity, detection limit, quantification limit, precision, accuracy, specificity, robustness and stability was accomplished in accordance with the International Conference on Harmonization (ICH) guidelines. The isocratic elution of fenofibrate was performed using Agilent 1260 Infinity HPLC system. The column was Capcell PAK C18 (4.6 mm x 250 mm, 5 μm). The mobile phase, consisted of acetonitrile and 0.1% (v/v) H3PO4 (75:25, v/v), was eluted at 2 ml/min. The eluent was monitored at 286 nm by the UV detector for fenofibrate concentration measurement. Stability test proved that fenofibrate in sample solutions remained stable at room temperature throughout the analytical process. The limit of detection (LOD) and limit of quantification (LOQ) furnished by this method were better than those of the previously reported HPLC and some UPLC methods of fenofibrate determination. All other validation parameters appeared within the acceptable limits.

Keywords: Acetonitrile, dosage form, fenofibrate, RP-HPLC, validation.

Purchase Online Rights and Permissions

  
  



Article Details

Volume: 10
Issue Number: 2
First Page: 97
Last Page: 104
Page Count: 8
DOI: 10.2174/1573412910999140113115903
Advertisement

Related Journals


Webmaster Contact: urooj@benthamscience.org Copyright © 2014 Bentham Science